This supplemental report is being submitted to provide the physical evaluation of the device and to update the following sections.A visual inspection was performed and found the distal rubber tip (where the guidewire is inserted) was missing and not returned.The missing piece is about 7 mm in length.The distal end of the device was closely examined and found minor scratch marks and a small portion of the rubber tip adhered inside the distal end.The v marking was bent and stressed.Also, there is a sticky foreign substance adhered to the pipe causing resistance while operating the pipe.The basket was found still intact as the basket could be opened, closed, and functioned without issue.Based on the investigation findings and similar reported events, the most probable cause of the reported event could be attributed to (unintended) operational error.The instruction manual provides warning which states, ¿when using the guide wire, insert the instrument with its distal tip in parallel with the guide wire.Careful not to forcibly insert the instrument with a sharp angle between the distal tip and the guide wire as this may damage the distal tip.Also, do not crush the calculus while the guidewire remains in the distal tip.This may also damage the instrument.¿.
|