Catalog Number VS-402
Device Problem Migration or Expulsion of Device (1395)
Patient Problem Patient Problem/Medical Problem (2688)
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
The physician uses a venaseal to treat one segment of the right short saphenous vein (ssv).Ifu was followed.Local anesthesia and hand compressions were performed.A guidewire was used for the insertion of the catheter and the segment was closed after the procedure.It was reported that upon a follow up ultrasound visit, the physician noted adhesive was located in the right popliteal vein.The patient was placed on a blood thinner for 30 days.
 
Manufacturer Narrative
Additional information: catheter was 5cm from the spj prior to delivery of adhesive.The physician speculates that, as the patient was in the trendelenburg position prior to treatment, this may have led to a bloodless field where the adhesive did not have enough blood to allow for polym erization.No device or photographic images were received for evaluation.A single photograph of a sonographic image was received.The image shows a thrombus formation in the right popliteal vein.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENA SEAL CLOSURE SYSTEM
Type of Device
AGENT, OCCLUDING, VASCULAR, PERMANENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
091708734
MDR Report Key8146598
MDR Text Key129778477
Report Number9612164-2018-03534
Device Sequence Number1
Product Code PJQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberVS-402
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/06/2018
Initial Date FDA Received12/10/2018
Supplement Dates Manufacturer Received12/17/2018
03/08/2019
Supplement Dates FDA Received01/15/2019
03/12/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age87 YR
-
-