Catalog Number VS-402 |
Device Problem
Migration or Expulsion of Device (1395)
|
Patient Problem
Patient Problem/Medical Problem (2688)
|
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The physician uses a venaseal to treat one segment of the right short saphenous vein (ssv).Ifu was followed.Local anesthesia and hand compressions were performed.A guidewire was used for the insertion of the catheter and the segment was closed after the procedure.It was reported that upon a follow up ultrasound visit, the physician noted adhesive was located in the right popliteal vein.The patient was placed on a blood thinner for 30 days.
|
|
Manufacturer Narrative
|
Additional information: catheter was 5cm from the spj prior to delivery of adhesive.The physician speculates that, as the patient was in the trendelenburg position prior to treatment, this may have led to a bloodless field where the adhesive did not have enough blood to allow for polym erization.No device or photographic images were received for evaluation.A single photograph of a sonographic image was received.The image shows a thrombus formation in the right popliteal vein.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|