Model Number E801 |
Device Problem
Non Reproducible Results (4029)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/08/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event occurred in nor.- attachment: [patient results mdr-29901.Pdf].
|
|
Event Description
|
The customer provided data for 4 patient samples tested on cobas e801 module serial number (b)(4).Based on the data provided the results for all 4 patient samples were discrepant when tested for folate, tsh or vitamin b12.Refer to attached data for the patient results.It is not known if any erroneous results were reported outside of the laboratory.There was no allegation that an adverse event occurred.No reagent lot numbers or expiration dates were provided.
|
|
Manufacturer Narrative
|
The root cause of the issue was related to defective assay tips.The tip lot used is covered by a roche initiated recall.Customer was informed to stop using the affected tips.
|
|
Manufacturer Narrative
|
The alarm trace does not contain any information relevant to the event.The investigation found the root cause was poor adjustment of the sampling area (sample probe rotation, tube gripper, and camera alignment).
|
|
Search Alerts/Recalls
|