Model Number EX1150
Device Problem Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/28/2018
Event Type  malfunction  
Manufacturer Narrative
The field complaint of the transmitter losing power after a lightning strike was verified.Final analysis found burnt damage on the green contact strips.The unit was not plugged into a working ac electrical outlet due to the burnt damage.Upon opening the transmitter, inspection revealed that there were multiple burnt components on the main pcb and the inner casing of the ac power adapter had a burnt mark.As a result it was concluded that during the lightning strike, the merlin@home transmitter was hit by a high current flow through the ac wall power outlet into the ac power adapter and through the telephone line into the transmitter, thus damaging their respective main pcb.
 
Event Description
This report is to advise of an event observed during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN@HOME, RF TELEMETRY BASIC
Type of Device
PACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8154063
MDR Text Key130024517
Report Number2017865-2018-18742
Device Sequence Number1
Product Code DRG
UDI-Device Identifier05414734504799
UDI-Public05414734504799
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial
Report Date 12/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberEX1150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/28/2018
Initial Date FDA Received12/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/09/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-