• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LIGHT SOURCE, 500XL, XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LIGHT SOURCE, 500XL, XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE Back to Search Results
Catalog Number 72200568
Device Problem Fire (1245)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/07/2018
Event Type  malfunction  
Event Description
It was reported that during a procedure, the power cord of light source 500xl burned during the operation.The place in which the fire started was at a part of a multi plug of power strip.The customer exchanged the power cord and re-connected with another power source, the device was successfully booted.The procedure was completed without patient injuries and complications.
 
Manufacturer Narrative
The reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A relationship, if any, between the subject device and the reported event could not be determined.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Manufacturer Narrative
The device was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of burned power cord was confirmed.The burnt power cord was not found shorted out and still functions.Light source passed functional testing.All functions perform as expected and light output was steady and continuous.The complaint was confirmed and the root cause was determined to be a defective power strip or an improper installation of the power cord into a multi plugged power strip causing an overloaded circuit.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGHT SOURCE, 500XL, XENON
Type of Device
IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 w. william cannon drive
austin, TX 78735
5123585706
MDR Report Key8154846
MDR Text Key130202459
Report Number1643264-2018-00977
Device Sequence Number1
Product Code FFS
UDI-Device Identifier03596010643070
UDI-Public03596010643070
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K994084
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 04/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200568
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2019
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/10/2018
Initial Date FDA Received12/12/2018
Supplement Dates Manufacturer Received02/05/2019
04/04/2019
Supplement Dates FDA Received02/06/2019
04/05/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-