The device was not returned for evaluation; it was discarded at the hospital.Without return of the unit it is not possible to determine if some damage or defect existed on the unit that could have contributed to the event.No corrective actions will be taken at this time.A review of the manufacturing records indicated that the product met specifications upon release.Poor dynamic response can be caused by air bubbles, clotting, and excessive lengths of tubing, excessively compliant pressure tubing, small bore tubing, loose connections, or leaks.The assembly may be tested for dynamic response by observing the pressure waveform on a monitor.Bedside determination of the dynamic response of the catheter, monitor, kit and transducer system is done after the system is flushed, attached to the patient, zeroed and calibrated.A square-wave test may be performed by pulling the snap tab device and releasing quickly.Pressure readings can change quickly and dramatically because of loss of proper calibration, loose connection, or air in the system.Pressure readings should correlate with the patient¿s clinical manifestations.There was no patient compromise noted in this event.It is unknown whether user or procedural factors played a role in this event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|