FISHER & PAYKEL HEALTHCARE LTD ICON SERIES CPAP; BZD Back to Search Results
Model Number ICONAAA
Device Problem Mechanical Problem (1384)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint icon cpap was not received at fisher & paykel healthcare (f&p) in (b)(4).Our investigation is accordingly based on the information and photograph provided by the customer.Results: visual inspection of the photograph revealed that the power cord had gross damage to the outer and inner insulation of the power cord, exposing copper wiring.Conclusion: the nature of the damage suggests the power cord may have been subjected to excessive force or sharp tools.The lot date confirms that the unit is almost two years old.During initial assembly of the icon cpap, all power cords are visually inspected for damage.Once the assembly process is completed, all icon cpap devices are visually inspected again before release for distribution.This suggests the damage occurred after it had been distributed.The icon cpap is designed to the electrical safety standards, ul60601-1 and as/nzs 3200.1.The materials used in the thermoplastic components of the mains connector and the cases are flame retardant according to ul 60601-1 and as/nzs 3200.1.Our user instructions that accompany the icon state the following: "only operate if the device, power cord and plug are dry and in good working order." "do not operate the device, water chamber or breathing tube if it is dropped, damaged or not working as intended.".
 
Event Description
A distributor in (b)(6) reported that an icon cpap humidifier had a damaged power cord.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICON SERIES CPAP
Type of Device
BZD
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology dr., suite 100
irvine, CA 92618
9494534000
MDR Report Key8158577
MDR Text Key130176795
Report Number9611451-2018-01032
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K094040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial
Report Date 11/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberICONAAA
Device Catalogue NumberICONAAA
Device Lot Number2100124363
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 11/13/2018
Initial Date FDA Received12/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/20/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-