Model Number 37800
Device Problem Energy Output Problem (1431)
Patient Problems Paresis (1998); Vomiting (2144); Therapeutic Response, Decreased (2271)
Event Date 12/10/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare professional (hcp) regarding a patient with an implantable neurostimulator (ins).It was reported that the patient was at the er and the hcp would like the device checked because the patient had a return of symptoms for 2 days.No further complications were reported/anticipated.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from a healthcare professional (hcp).It was reported that the patient was not seen in the local er and the changes were not done by the hcp.It was noted that the cause of the patient¿s return of symptoms was gastroparesis.There was a change in cycle time, and an increase of domperidone back to 3 a day.No further complications were reported/anticipated.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from a consumer.It was reported that the patient had been throwing up yellow bile-looking stuff and could not seem to get it to stop.The healthcare provider (hcp) called a manufacturer representative and talked to them about it; a ¿tester¿ was overnighted to the hospital where the patient went and they checked the device and increased the voltage as the manufacturer representative suggested.It was noted that the patient¿s symptoms then improved, but when the patient went to the hcp, the hcp brought the voltage back down and upped the cycle, and ever since then, the patient was not doing as well.The patient was redirected to follow up with their hcp regarding their symptoms.No further complications were reported/anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key8160575
MDR Text Key130684558
Report Number3004209178-2018-27580
Device Sequence Number1
Product Code LNQ
UDI-Device Identifier00643169614246
UDI-Public00643169614246
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 03/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/28/2018
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/12/2018
Initial Date FDA Received12/13/2018
Supplement Dates Manufacturer Received01/21/2019
02/11/2019
Supplement Dates FDA Received01/31/2019
03/05/2019
Date Device Manufactured09/29/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age62 YR
Patient Weight48
-
-