• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GETINGE (SUZHOU) CO., LTD. AUTO LOGIC; MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GETINGE (SUZHOU) CO., LTD. AUTO LOGIC; MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE Back to Search Results
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Unclear Information (4052)
Patient Problems Fall (1848); No Consequences Or Impact To Patient (2199)
Event Date 11/19/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Additional information will be provided upon the investigation conclusion.
 
Event Description
Arjo was notified by a customer that a patient fell from the bed while was sleep.It was not indicated that the safety sides were raised or not.
 
Manufacturer Narrative
Currently, we are still in the process of gathering additional information regarding the circumstances of the reported event.Further information will be provided upon conclusion of the investigation.
 
Manufacturer Narrative
Investigation has been carried out onto this incident and conclusions are following.Arjo was informed about a patient fall from arjo auto logic mattress and non-arjo bed frame (brand name was not provided).It is unknown if safety sides were in raised position or lowered.Specific patient medical condition was not provided.No allegation of product malfunction was reported by the customer.The product user guide instruction for use (ifu) number (b)(4) warns: "it is responsibility of the care giver to ensure that the user can use this product safely." "whilst the patient is unattended, the decision to use safety sides should be based on clinical assessment and in line with local policy".When reviewing reportable complaints for patient fall, we have found some incidents of patient fall, and 10 incidents have been reported by the one group of facilities from french market.After the events, arjo takes this matter very seriously and this is why arjo quality director for french market contacted the customer to help working on identifying the root cause of the issue.The customer facility representative stated that they are considering assessing patients on admit in order to select air mattresses for individual patient's needs.In conclusion, the arjo auto logic system (both mattress and pump) was used while the event occurred, therefore it played role in the event, however there was no allegation of device malfunction.The circumstances of the event, patient medical condition or bed frame used are unknown.Therefore, it is not possible to determine the exact root cause of patient fall.We report this event due to allegation of patient fall, which may lead to serious health consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTO LOGIC
Type of Device
MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE
Manufacturer (Section D)
GETINGE (SUZHOU) CO., LTD.
no. 158 fangzhou road,
jiangsu, suzhou 21502 4
CH  215024
MDR Report Key8163686
MDR Text Key130314383
Report Number3005619970-2018-00018
Device Sequence Number1
Product Code FNM
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup,Followup
Report Date 01/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/31/2019
Distributor Facility Aware Date11/19/2018
Event Location Hospital
Date Report to Manufacturer01/31/2019
Initial Date Manufacturer Received 11/19/2018
Initial Date FDA Received12/14/2018
Supplement Dates Manufacturer Received11/19/2018
11/19/2018
Supplement Dates FDA Received01/10/2019
01/31/2019
Patient Sequence Number1
Patient Outcome(s) Other;
-
-