• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FT4 III ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FT4 III ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 07976887190
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/27/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).
 
Event Description
The customer complained of questionable thyroid results for 3 patient samples on a cobas 8000 e 801 module.The patient samples were submitted for investigation.There were discrepant results identified for elecsys tsh, elecsys ft4 iii, and elecsys ft3 iii assay between the customer's e801 module, an e801 module used at the investigation site, a cobas 8000 e 602 module used at the investigation site, a cobas e 411 immunoassay analyzer used at the investigation site, and the centaur method.It was not known if the erroneous results from the customer site were reported outside of the laboratory.This medwatch will cover ft4 iii.Refer to medwatch with patient identifier (b)(6) for information on the tsh results, and medwatch with patient identifier (b)(6) for information on the ft3 iii results.There was no allegation of an adverse event.The customer's cobas e801 serial number was not provided.The serial number for the cobas e801 used at the investigation site was (b)(4).The tsh reagent lot used on this cobas e801 was 283556 with an expiration date of 28-feb-2019.The ft4 iii reagent lot used on this cobas e801 was 309694 with an expiration date of 31-jan-2019.The ft3 iii reagent lot used on this cobas e801 was 314824 with an expiration date of 30-may-2019.The serial number for the cobas e602 used at the investigation site was (b)(4).The serial number for the cobas e411 used at the investigation site was (b)(4).The tsh reagent lot used on the cobas e602 and the cobas e411 was 341699 with an expiration date of 28-feb-2019.The ft4 iii reagent lot used on the cobas e602 and the cobas e411 was 331797 with an expiration date of 31-may-2019.The ft3 iii reagent lot used on the cobas e602 and the cobas e411 was 314666 with an expiration date of 28-feb-2019.The investigation is ongoing.
 
Manufacturer Narrative
The customer¿s results were replicated and no interfering factors could be detected.Differences in results and reference ranges by different manufactures, in this case siemens, can vary due to differences in the types of antibodies used, setups of the assays, and standardization methodologies.The investigation did not identify a general reagent or product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III ASSAY
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key8163723
MDR Text Key130787689
Report Number1823260-2018-04860
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 02/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07976887190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 11/27/2018
Initial Date FDA Received12/14/2018
Supplement Dates Manufacturer Received11/27/2018
Supplement Dates FDA Received02/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-