MERIT MEDICAL SINGAPORE PTE LTD. SAFEDRAW BLOOD SUPPLY SYSTEM Back to Search Results
Catalog Number 686304/B
Device Problem Fluid/Blood Leak (1250)
Patient Problem Blood Loss (2597)
Event Date 11/15/2018
Event Type  malfunction  
Manufacturer Narrative
The suspect device is not returning for evaluation.The complaint could not be confirmed.The root cause could not be determined.A review of the device history and complaint database could not be performed since the lot number was not provided.
 
Event Description
The account alleges that a leak was identified near the arterial line connection.The device was replaced.No patient injury to report.
 
Manufacturer Narrative
The suspect device was returned for evaluation.The complaint was originally reported as a leak at the connector, however, visual examination of the returned device found the connector to be detached.The root cause is attributed to the manufacturing process.A review of the device history record was performed and no exception documents were found.A review of the complaint database was performed and no similar complaints for this lot number were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFEDRAW BLOOD SUPPLY SYSTEM
Type of Device
BLOOD SUPPLY SYSTEM
Manufacturer (Section D)
MERIT MEDICAL SINGAPORE PTE LTD.
198 yishun ave 7
singapore, 76892 6
SN  768926
MDR Report Key8164209
MDR Text Key130348602
Report Number8020616-2018-00083
Device Sequence Number1
Product Code DSK
Combination Product (y/n)N
PMA/PMN Number
K885235
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number686304/B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2019
Initial Date Manufacturer Received 11/16/2018
Initial Date FDA Received12/14/2018
Supplement Dates Manufacturer Received02/27/2019
Supplement Dates FDA Received03/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-