• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+ Back to Search Results
Catalog Number 10379675
Device Problem False Negative Result (1225)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/29/2018
Event Type  malfunction  
Manufacturer Narrative
Siemens engineer evaluated the instrument and did not find any hardware issues by running check cassette five times along with electronics and test table testing.The qc results were in the expected ranges.A successful correlation study was run with fresh positive and negative urine samples.The cause of this event is unknown.
 
Event Description
The customer reported a false negative and borderline urine hcg result from the clinitek status+ compared to a positive result from a home pregnancy test and from a non-siemens qualitative manual test kit.The customer stated that the patient is believed to be pregnant although there was no confirmatory serum test run.There was no report of injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK STATUS+
Type of Device
CLINITEK STATUS+
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury, CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
tim krauskopf
2 edgewater drive
norwood, MA 02062
8622285388
MDR Report Key8164902
MDR Text Key130504482
Report Number3002637618-2018-00135
Device Sequence Number1
Product Code JIL
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K091216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10379675
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/29/2018
Initial Date FDA Received12/14/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-