Maquet medical systems,usa(importer)submits this report on behalf of the legal manufacturer of the device maquet cardiopulmonary (b)(4).Rastatt, germany.A follow-up medwatch will be submitted when additional information becomes available.Reference exemption # (b)(4).Importer- maquet medical systems usa (b)(4).Contact person- (b)(6).Investigation: service observed the system and determined the motor and-or cam assembly was in need of replacement.During troubleshooting the belts were replaced due to broken belt.The device was returned to national repair center mahwah.According to the service report, the service team in mahwah found a worn tacho strobe, which has to be replaced.Thus the failure could be confirmed.Root cause could be determined as worn tacho strobe.As stated by the ssu technician the defective parts are no longer available for return.Since the parts are not available for return, further investigation has not been performed.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further iinvestigation initiations will be completed at this time.Correction: according to the service report the worn tacho strobe (701007785 mcp00923311#tacho strobe rpm) has been replaced.Pm, calibration and full functional test as per the service manual were performed.All tests passed.Corrective action: since the reported failure did not contribute to a death or serious injury no corrective action is needed.In addition at this time it cannot be concluded that this is a systemic error.No corrective action is needed.
|