• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING Back to Search Results
Model Number 3851
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/29/2018
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.
 
Event Description
It was reported that a balloon rupture occurred.The 90% stenosed target lesion was located in the moderately tortuous and severely calcified left anterior descending artery (lad).A 10mm x 2.00mm wolverine coronary cutting balloon was used to dilate the lesion and on the first inflation at 6 atmospheres, the balloon ruptured.The procedure was completed with a non bsc device.No patient complications were reported in relation to this event.
 
Event Description
It was reported that a balloon rupture occurred.The 90% stenosed target lesion was located in the moderately tortuous and severely calcified left anterior descending artery (lad).A 10mm x 2.00mm wolverine coronary cutting balloon was used to dilate the lesion and on the first inflation at 6 atmospheres, the balloon ruptured.The procedure was completed with a non bsc device.No patient complications were reported in relation to this event.
 
Manufacturer Narrative
Age at time of event: 18 years or older.Device evaluated by mfr: a visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Blood was identified within the balloon and inflation lumen which is evidence of a device leak.The returned device was attached to a boston scientific encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located approximately 5mm distal to the distal end of the proximal markerband.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.The rated burst pressure for this device is 12 atmospheres.The markerbands, blades and tip section of the device were visually and microscopically examined, and no issues were noted with the device that may have potentially contributed to the complaint incident.All blades were present and fully bonded to the balloon material.A visual and tactile examination found no damage or issues with the shaft or hypotube of the device.No other issues were identified during the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON MONORAIL
Type of Device
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8171400
MDR Text Key130631541
Report Number2134265-2018-63908
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 07/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/07/2020
Device Model Number3851
Device Catalogue Number3851
Device Lot Number0022763357
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2019
Initial Date Manufacturer Received 11/29/2018
Initial Date FDA Received12/17/2018
Supplement Dates Manufacturer Received06/12/2019
Supplement Dates FDA Received07/02/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-