Reportable based on analysis completed 12 december 2018.It was reported that the comet pressure guidewire would not unlock from the handle.There were no patient issues related to this event, however, the returned device analysis revealed some peeling of the device coating.The returned product consisted of a ffr comet pressure wire connected to the rfid cable.The tip, device shaft, sensor port and the coefficient values were examined for damage or any irregularities.The shaft showed three kinks located on the shaft.The first kink was located 92.5 cm from the tip.The second kink was located 152.5cm from the tip and the third kink 162.5cm from the tip.There was some peeling of the coating at the 152.5cm location.The rotating handle cap was loosened and the wire was removed from the handle.There was no issue in removing the wire.The wire was reinserted and the sequence was repeated with no issues.The comet wire was then connected to the ffr link for signal verification.The signal was present as designed.The sensor port showed no residue of body fluids.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The pressure wire was connected to the analysis support test bench and all applicable data was correct as designed, there was no difficulty in connecting the wire.The coefficient values were confirmed to be in specification.
|