The cause for the (b)(6) advia centaur xp anti-hbs2 (ahbs2) patient result and out of range quality control (qc) results with the new reagent lot (119096) is unknown.There has been no similar issue with other patient samples during this occurrence.Repeat testing with (b)(6) on (b)(6) 2018 showed consistent result to lot 119090 as a (b)(6) result ((b)(6)) was obtained with lot 119096.As part of troubleshooting testing, anti-hbs2 master curve material (mcm, lot 71943) showed acceptable results between lot 119096 and lot 119102.Only mcm level 5 was evaluated with reagent lot 119090 due to sample availability during testing.Additional testing with patient samples showed similar results between lot 119090 and lot 119096.The customer accepted lot 119096 as per laboratory protocol, and adjusted qc ranges accordingly.Qc results were similar to bio-rad peer data for virotrol ii qc.A review of in-house data showed the performance of reagent lot 119096 met all release criteria.Master curve material (mcm) results (miu/ml):(b)(6).Additional patient samples run on reagent lot 119096 had similar results to reagent lot 119090 (actual results not provided).The instrument is performing within specification.No further evaluation of the device is required.
|