MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S3 ERC TUBING CLAMP/ 620MM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S3 ERC TUBING CLAMP/ 620MM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-05-00
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Patient Involvement (2645)
Event Date 11/19/2018
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s3 erc tubing clamp/ 620mm.The incident occurred in (b)(6).Livanova (b)(4) requested the part back for further investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a s3 erc tubing clamp displayed error messages during servicing.There was no patient involvement.
 
Manufacturer Narrative
The defective device was returned to livanova deutschland for further investigation.During the evaluation the reported issue could be reproduced.As root cause, fluid ingress inside the device could be determined.Livanova deutschland implemented a corrective action for the reported issue.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.
 
Event Description
See initial.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S3 ERC TUBING CLAMP/ 620MM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8178172
MDR Text Key131000319
Report Number9611109-2018-01411
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K950990
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-05-00
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/19/2018
Initial Date FDA Received12/19/2018
Supplement Dates Manufacturer Received01/09/2019
Supplement Dates FDA Received02/07/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-