MAQUET CRITICAL CARE AB FLOW-I C30; GAS-MACHINE, ANESTHESIA Back to Search Results
Model Number C30
Device Problem High Readings (2459)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/17/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the anesthesia workstation measured a higher fico2 level than expected during an extended time period.There was no patient harm.(b)(4).
 
Manufacturer Narrative
Getinge usa sales, llc (importer) is submitting this report on behalf of maquet critical care ab (manufacturer).Ref.Exemption #:(b)(4).Getinge usa sales, llc (b)(4).Contact peron: (b)(4).The investigation into this matter consists of an evaluation of the received device logs only as no parts have been reported as replaced.The received logs confirm the reported issue with high fico2 readings.A successful sco was performed prior to the event indicating that no technical malfunctions were present.Alarms for high fico2 level were generated during ongoing patient treatment.The logs show further that the co2 absorber was replaced during the case and that the user also tried to rectify the issue by increasing the fresh gas flow and activate the o2 flush several times without success.The received trend log confirms the high fio2 readings.We have not been able to determine the true cause of the reported issues.
 
Event Description
Manufacturer's ref #:(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOW-I C30
Type of Device
GAS-MACHINE, ANESTHESIA
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
solna
SW 
MDR Report Key8178681
MDR Text Key131138121
Report Number8010042-2018-00680
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K160665
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC30
Device Catalogue Number6677300
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/18/2018
Initial Date FDA Received12/19/2018
Supplement Dates Manufacturer Received05/14/2019
Supplement Dates FDA Received05/14/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-