Catalog Number PLEE60A
Device Problems Failure to Advance (2524); Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/26/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.As a lot/batch was not provided, a device history could not be performed.
 
Event Description
It was reported that during a laparoscopic sleeve gastrectomy, while using seam guard the device would not pass through the trocar easily."it appeared that the anvil was bent and the distal tip would not come together".The procedure was completed with this device and there were no patient consequences reported.
 
Manufacturer Narrative
(b)(4).Batch # r5a98g.Investigation summary: the analysis results found that one plee60a device was returned with the anvil bent upwards and with a gst60g reload loaded on the device.The reload was received fully fired.No functional test was performed due the condition.It is possible that the device was clamped over an excess of tissue causing the anvil to bend and for the firing stroke and the staple form to be incomplete.It should be noted that a 100% inspection takes place during manufacturing to ensure the device meets the required specifications prior to shipments, in addition, a sample of the batch is inspected at fgqa.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERED 60 ECHELON +, 440MM SHAFT
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key8179710
MDR Text Key131035352
Report Number3005075853-2018-15266
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10705036014614
UDI-Public10705036014614
Combination Product (y/n)N
PMA/PMN Number
K110385
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 11/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/08/2021
Device Catalogue NumberPLEE60A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2019
Initial Date Manufacturer Received 11/26/2018
Initial Date FDA Received12/19/2018
Supplement Dates Manufacturer Received05/03/2019
Supplement Dates FDA Received05/10/2019
Patient Sequence Number1
-
-