Catalog Number PLEE60A |
Device Problems
Failure to Advance (2524); Material Twisted/Bent (2981)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/26/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.As a lot/batch was not provided, a device history could not be performed.
|
|
Event Description
|
It was reported that during a laparoscopic sleeve gastrectomy, while using seam guard the device would not pass through the trocar easily."it appeared that the anvil was bent and the distal tip would not come together".The procedure was completed with this device and there were no patient consequences reported.
|
|
Manufacturer Narrative
|
(b)(4).Batch # r5a98g.Investigation summary: the analysis results found that one plee60a device was returned with the anvil bent upwards and with a gst60g reload loaded on the device.The reload was received fully fired.No functional test was performed due the condition.It is possible that the device was clamped over an excess of tissue causing the anvil to bend and for the firing stroke and the staple form to be incomplete.It should be noted that a 100% inspection takes place during manufacturing to ensure the device meets the required specifications prior to shipments, in addition, a sample of the batch is inspected at fgqa.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
|