• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. XPS® BLADE; BUR, EAR, NOSE AND THROAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. XPS® BLADE; BUR, EAR, NOSE AND THROAT Back to Search Results
Model Number 1882040HR
Device Problem Material Separation (1562)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/22/2018
Event Type  malfunction  
Manufacturer Narrative
Analysis found that there were only trace amounts of a residue consistent with biological contaminants on the device.The tip was present and in good condition.There were no bends or concentricity issues.There were no loose components.There was deformation of the front hub.Functionally, the blade loaded into a handpiece ran at 3,000 rpm in oscillate mode, and cut saw bone with no issues.If information is provided in the future, a supplemental report will be issued.
 
Event Description
A health care provider (hcp) reported via a manufacturer representative that the tip of the blade broke during an endoscopic sinus surgery procedure.The hcp noted that the inner cylinder of the blade was broken, not the external cylinder.There were no broken pieces remained in the patient's body.There was no intervention planned or performed as a result of this event.There was 10 minutes procedure delay.The procedure was completed with a back up device.There was no patient impact or injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® BLADE
Type of Device
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
uriza shums
6743 southpoint drive north
jacksonville, FL 32216
9043328405
MDR Report Key8181301
MDR Text Key130981650
Report Number1045254-2018-00722
Device Sequence Number1
Product Code EQJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1882040HR
Device Catalogue Number1882040HR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/22/2018
Initial Date FDA Received12/19/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-