• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM5_12.6
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Dizziness (2194); No Code Available (3191)
Event Date 07/12/2018
Event Type  Injury  
Manufacturer Narrative
This product is not marketed in the us.(b)(4).Work order search: no similar complaint type events were reported for units within the same lot.(b)(4).
 
Event Description
The reporter indicated that the surgeon implanted a 12.6mm, vticm5_12.6, -9.00/+2.0/95 (sphere/cylinder/axis), implantable collamer lens into the patients right eye (od) on (b)(6) 2018.The patient experienced dizziness and feeling seasickness after initial surgery.It was reported that it was not refractive surprise.On (b)(6) 2018 the lens was exchanged for a non toric lens.This exchange resolved the problem.Cause of event was unknown.
 
Manufacturer Narrative
Device evaluation: the lens was returned in liquid in a micro centrifuge vial.Visual inspection found no visible damage to lens.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
MDR Report Key8182707
MDR Text Key130976279
Report Number2023826-2018-01762
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2021
Device Model NumberVTICM5_12.6
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/21/2018
Initial Date FDA Received12/19/2018
Supplement Dates Manufacturer Received01/15/2019
Supplement Dates FDA Received01/28/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/16/2018
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age33 YR
-
-