Model Number 816571 |
Device Problem
Mechanical Jam (2983)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
As per the perfusionist, this event occurred over a month ago.The field service representative (fsr) was only able to duplicate the reported complaint but over occluding the pump.The unit operated to the manufacturer's specifications.
|
|
Event Description
|
It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, there were multiple instances where the pump being used in the cardioplegia (cpg) position jammed.The outlet was clamped, tubing was readjusted, and the flow was set at a lower rate.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Event Description
|
Additional information was received that the reported issue occurred during set-up.
|
|
Manufacturer Narrative
|
Per data log analysis, on (b)(6) 2018 the pump did report underspeed (head < demand) three times.This could explain the reported issue while the pump was still running.The complaint does not indicate the date the problem occurred.
|
|
Search Alerts/Recalls
|