• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090
Device Problem Device Emits Odor (1425)
Patient Problem No Patient Involvement (2645)
Event Date 12/06/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the programmer had a burning smell when turned on.The programmer was returned for service.There was no patient involvement.
 
Manufacturer Narrative
Product analysis: manufacturer's analysis confirmed the customer comment that the programmer had a burning smell when turned on.It was indicated that the power supply was out of specification.The programmer was repaired and returned to use.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key8184552
MDR Text Key131013198
Report Number2182208-2018-02298
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00613994175397
UDI-Public00613994175397
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 01/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2018
Initial Date Manufacturer Received 12/06/2018
Initial Date FDA Received12/20/2018
Supplement Dates Manufacturer Received01/07/2019
Supplement Dates FDA Received01/10/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/26/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-0406-2019
Patient Sequence Number1
Treatment
2067L RADIOFREQUENCY HEAD, 229047 ANALYZER
-
-