• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO BROCKENBROUGH NEEDLE; TROCAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO BROCKENBROUGH NEEDLE; TROCAR Back to Search Results
Model Number EP003994S
Device Problems Break (1069); Activation, Positioning or Separation Problem (2906)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/23/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a radiofrequency (rf) procedure, skiving occurred when the needle scratched in the inside of the sheath.The sheath was replaced with resolve.The case was completed with rf.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the transseptal needle, (b)(4) with lot number 215903098, was not returned.While no product was returned, the oem evaluated the sheath used and determined that a combination of the products is potentially susceptible to skiving.The cause of the issue was not established.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BROCKENBROUGH NEEDLE
Type of Device
TROCAR
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key8185969
MDR Text Key131139356
Report Number9612164-2018-03747
Device Sequence Number1
Product Code DRC
UDI-Device Identifier00613994663146
UDI-Public00613994663146
Combination Product (y/n)N
Reporter Country CodeGR
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/25/2020
Device Model NumberEP003994S
Device Catalogue NumberEP003994S
Device Lot Number215903098
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/13/2018
Initial Date FDA Received12/20/2018
Supplement Dates Manufacturer Received03/06/2020
Supplement Dates FDA Received03/16/2020
Date Device Manufactured07/26/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-