• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PATELLO-FEMORAL TROCHLEA COMPONENT PRECOAT; PROSTHESIS KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PATELLO-FEMORAL TROCHLEA COMPONENT PRECOAT; PROSTHESIS KNEE Back to Search Results
Model Number N/A
Device Problem Migration or Expulsion of Device (1395)
Patient Problems Pain (1994); Swelling (2091)
Event Date 12/04/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: catalog #: 00584201302, femoral component high flex precoat, lot # 62827179.Catalog #: 00584202408, articular surface size 4, lot # 62697221.Catalog #: 00584200402, tibial component precoat, lot # 62828687.Customer has indicated that the product will not be returned to zimmer biomet for investigation, due to being discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event; please see associated reports: 0001822565-2018-07028.
 
Event Description
It was reported that a patient underwent an initial knee procedure on approximately 3 years ago.Subsequently, the patient was revised due to pain, loosening, and instability.
 
Manufacturer Narrative
(b)(4).Reported event was considered to be confirmed as the reported harms were noted in the medical records.The x-ray also noted loosening and possible joint effusion.Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.There are warnings in the package insert that this type of event can occur and risks are addressed in risk documentation.A summary of the investigation was sent to the complainant conveying proper surgical technique and use of the device.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATELLO-FEMORAL TROCHLEA COMPONENT PRECOAT
Type of Device
PROSTHESIS KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8186449
MDR Text Key131070282
Report Number0001822565-2018-07029
Device Sequence Number1
Product Code KRR
Combination Product (y/n)N
PMA/PMN Number
K070695
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00592601102
Device Lot Number62936327
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/03/2018
Initial Date FDA Received12/20/2018
Supplement Dates Manufacturer Received09/18/2019
Supplement Dates FDA Received09/23/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Weight59
-
-