MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number ATG120164
Device Problems Peeled/Delaminated (1454); Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the device is pending.The investigation of the reported event is currently underway.
 
Event Description
It was reported that during an iliofemoral angioplasty procedure, the pta balloon allegedly ruptured on the initial inflation attempt.There was no reported patient injury.
 
Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was returned for evaluation.A visual inspection found a peeled outer layer on the distal cone of the balloon.An inflation attempt was made and a pinhole rupture was noted on the distal cone at the location of the peeling outer layer.Therefore, the investigation is confirmed for a pinhole rupture and peeled outer layer.It is likely that the rupture contributed to the peeling outer layer.However, the definitive root cause for the identified issues could not be determined based on the available information.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
 
Event Description
It was reported that during an iliofemoral angioplasty procedure, the pta balloon allegedly ruptured on the initial inflation attempt.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS GOLD PTA DILATATION CATHETER
Type of Device
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8189383
MDR Text Key131164128
Report Number2020394-2018-02243
Device Sequence Number1
Product Code LIT
UDI-Device Identifier00801741060823
UDI-Public(01)00801741060823
Combination Product (y/n)N
PMA/PMN Number
K181323
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Type of Report Initial,Followup
Report Date 05/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberATG120164
Device Catalogue NumberATG120164
Device Lot Number93UC0095
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2019
Initial Date Manufacturer Received 11/30/2018
Initial Date FDA Received12/21/2018
Supplement Dates Manufacturer Received05/10/2019
Supplement Dates FDA Received05/13/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-