Catalog Number 391350 |
Device Problem
Migration or Expulsion of Device (1395)
|
Patient Problems
Extravasation (1842); Necrosis (1971)
|
Event Date 11/29/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(6).A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
|
|
Event Description
|
It was reported that while using bd neoflon¿ iv cannula there were 2 extravasations in baby's.There was skin necrosis and the babies needed surgery.
|
|
Manufacturer Narrative
|
Investigation summary: a review of the device history record revealed no irregularities during the manufacture of the reported lot.1 representative sample (batch #8052239) was returned for investigation.The 1 representative sample was subjected to visual inspection, tip od measurement and bevel angle measurement.The 1 representative sample passed the acceptance criteria.The root cause cannot be determined as the 1 representative sample passed the acceptance criteria.Based on the investigation of previous similar complaint, the probable root cause could be due to the tubing material.Capa #81917 was issued to review the tubing material.
|
|
Event Description
|
It was reported that while using bd neoflon¿ iv cannula there were 2 extravasations in baby's.There was skin necrosis and the babies needed surgery.
|
|
Search Alerts/Recalls
|