Catalog Number 512535VER
Device Problems Fracture (1260); Material Separation (1562)
Patient Problem Foreign Body In Patient (2687)
Event Date 12/11/2018
Event Type  Injury  
Manufacturer Narrative
The device is not expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event Description
The account alleges that during a chronic total occlusion [cto] procedure, the 5f catheter tip detached within the patient.The physician had acquired right retrograde femoral artery access and had negotiated the patient's aortic bifurcation in order to cannulate the left superficial femoral artery [sfa].While positioning the catheter resistance was experienced due to a previously placed peripheral vascular stent and the catheter tip detached within the patients sfa.The physician used a vascular snare device to successfully remove the catheter tip from the patient.
 
Manufacturer Narrative
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A review of the device history record was performed and no exception documents were found.A review of the complaint database was performed and no similar complaints for this lot number were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPRESS PERIPHERAL CATHETER
Type of Device
PERIPHERAL CATHETER
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS, INC.
14646 kirby drive
houston TX 77047
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS, INC.
14646 kirby drive
houston TX 77047
Manufacturer Contact
katie swenson cqe,cba,cqpa.
1600 merit parkway
south jordan, UT 84095
8012531600
MDR Report Key8190096
MDR Text Key131173527
Report Number3010665433-2018-00080
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K093004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2021
Device Catalogue Number512535VER
Device Lot NumberE1383155
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/13/2018
Initial Date FDA Received12/21/2018
Supplement Dates Manufacturer Received02/01/2019
Supplement Dates FDA Received02/08/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/07/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-