A patient had a prophylactic full revision surgery, and the explanted devices were returned for analysis.Analysis of the returned lead noted abrasion on the outer silicone tubing at multiple locations.It was also noted that there was an abraded opening identified in the inner silicone tubing of the negative coil.Dried remnants of what appear to have once been body fluids/betadine solution were located inside the inner and the outer silicone tubing due to points of entrance at abraded openings and the cut ends of the returned lead portions.Other than the inner/outer tubing abrasions and typical wear and explant related observations, no other anomalies were identified in the returned lead portions.Analysis of the returned generator identified no anomalies.No additional relevant information was received to date.
|