• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL INFANT FLOW PLUS INFANT CPAP SYSTEM WITH SIPAP FUNCTION; VENTILATOR, NON-CONTINUOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL INFANT FLOW PLUS INFANT CPAP SYSTEM WITH SIPAP FUNCTION; VENTILATOR, NON-CONTINUOUS Back to Search Results
Model Number SIPAP
Device Problem Mechanical Problem (1384)
Patient Problem No Patient Involvement (2645)
Event Date 11/29/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Vyaire medical was able to confirm the customer's reported issue during testing and evaluation.This reported issue was due to a defective flow meter.Vyaire medical replaced the flow meter to address the customer's reported issue and performed the 2 year preventative maintenance on the unit.
 
Event Description
It was reported to vyaire medical that the flow meter was erratic.There was no report of any patient involvement associated with this reported event.
 
Manufacturer Narrative
Results of investigation: vyaire medical conducted failure analysis testing on the defective component.Visual inspection of the assembly revealed no signs of any physical damages.The flow meter was installed into a test fixture and powered on to service mode.The test fixture was checked for errors on the screen, no errors were presented.The user verification test was performed and the leak test.The unit passed by confirming the setting.Pressure calibrations were also performed, and all calibrations were passed; cleared all faults by pressing error button, no anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFANT FLOW PLUS INFANT CPAP SYSTEM WITH SIPAP FUNCTION
Type of Device
VENTILATOR, NON-CONTINUOUS
Manufacturer (Section D)
VYAIRE MEDICAL
22745 savi ranch pkwy
yorba linda CA 92887
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
stanley tan
22745 savi ranch pkwy
yorba linda, CA 92887
7149193324
MDR Report Key8191960
MDR Text Key131307268
Report Number2021710-2018-09714
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031745
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSIPAP
Device Catalogue Number27416-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/29/2018
Initial Date FDA Received12/21/2018
Supplement Dates Manufacturer Received04/12/2019
Supplement Dates FDA Received06/28/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/30/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-