Model Number 9735602 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/28/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device lot number, or serial number, unavailable.Manufacturing date was unavailable at the time of filing.No parts have been returned to the manufacturer for analysis.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a navigation device being used for a functional endoscopic sinus surgery (fess).It was reported that the magnetic field stopped responding in the middle of using the reported product.When checking the setting screen, it was found that the line of the emitter became orange and the magnetic field failed to respond.No improvement was observed despite taking actions such as disconnecting and reconnecting the cord.The system was rebooted and the issue was resolved.There was a reported delay of 5 minutes and no reported impact to patient outcome.
|
|
Manufacturer Narrative
|
A medtronic representative went to the site to test the equipment.It was reported that the issue could not be confirmed or replicated, no components were replaced.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: the system serial number was received and added.If information is provided in the future, a supplemental report will be issued.
|
|