MAUDE Adverse Event Report: SIRONA DENTAL SYSTEMS GMBH MIDWEST E PLUS 1:5 HIGHSPEED ATTACHMENT; HANDPIECE, BELT AND/OR GEAR DRIVEN, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIRONA DENTAL SYSTEMS GMBH MIDWEST E PLUS 1:5 HIGHSPEED ATTACHMENT; HANDPIECE, BELT AND/OR GEAR DRIVEN, DENTAL Back to Search Results
Catalog Number 875205
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
While no serious injury resulted in this event, there has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
It was reported that a midwest e plus 1:5 ca overheated during use; no injury resulted.
 
Manufacturer Narrative
Handpiece didn't get hot.Handpiece failed specs due to cap sticking inwards position when pressing on the cap causing contact with rotor.Cap is a new style cap.Cap is clean and no signs of debris around cap.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDWEST E PLUS 1:5 HIGHSPEED ATTACHMENT
Type of Device
HANDPIECE, BELT AND/OR GEAR DRIVEN, DENTAL
Manufacturer (Section D)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, D-646 25
GM  D-64625
Manufacturer (Section G)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM   D-64625
Manufacturer Contact
karl nittinger
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494424
MDR Report Key8196274
MDR Text Key131505756
Report Number9614977-2018-00207
Device Sequence Number1
Product Code EFA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K965238
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 01/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number875205
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/19/2018
Initial Date FDA Received12/26/2018
Supplement Dates Manufacturer Received12/14/2018
Supplement Dates FDA Received01/04/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-