The device was returned to omsc for evaluation.During the evaluation, omsc confirmed the reported event and a pinhole at the instrument channel of the device.Furthermore, it was confirmed that the inside of the control section of the device significantly corroded.No damage was found on the bending section and the angulation wire of the device.Omsc reviewed the manufacturing history of the device and confirmed no irregularity.Omsc concluded that the cause of the reported event was the corrosion in the control section of the device.The corrosion was highly likely caused by solution, which was used during reprocessing of procedures and invaded through the pinhole portion of the instrument channel.Based on the investigation result of the inner surface of the instrument channel, the pinhole possibly occurred due to an insertion of endo therapy accessory by excessive force.
|