• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME Back to Search Results
Model Number N/A
Device Problem Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/27/2018
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Report source - foreign - (b)(6).The investigation is still in progress.Once the investigation is complete a follow up mdr will be submitted.Not returning.
 
Event Description
It was reported that the headpiece of the unit was losing power.No adverse events were reported as a result of this malfunction.
 
Event Description
No additional information.
 
Manufacturer Narrative
This follow up report is being submitted to report additional information.On december 3, 2018, it was reported that the unit was losing power.The customer returned an air dermatome device, serial number (b)(4), for evaluation.The device history record (dhr) review noted no related non-conformances, requests for deviation, change notices or any other issues with manufacturing or with the device.The dhr review also found that all verifications, inspections and tests were successfully completed.Zimmer biomet australia has not previously repaired/evaluated air dermatome serial number (b)(4) as documented in the repair reports in livelink.Product review of the air dermatome by zimmer biomet australia on december 3, 2018 revealed that the head and depth bar had visible signs of wear.Product review of the air dermatome by zimmer biomet taiwan on march 27, 2019 revealed that the device did not operate when tested.Repair of the air dermatome was performed by zimmer biomet taiwan on march 29, 2019 which included replacement of the motor, machined head, bearings, o-ring, seal, reciprocating arm, external e-ring, lever, ball plunger and width plate screws.Air dermatome, serial number (b)(4), was then tested and functioned properly.It was repaired, inspected and tested.The reported event was confirmed since during the product review by zimmer biomet taiwan it was noted that the device did not operate when tested.The root cause of the reported event could not be specifically determined with the information that was provided.During the product review by zimmer biomet taiwan it was noted that the device did not operate when tested.It is unknown with the information that was provided how this occurred.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
MDR Report Key8200447
MDR Text Key131725250
Report Number0001526350-2018-01139
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 04/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number2765200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/03/2018
Initial Date FDA Received12/27/2018
Supplement Dates Manufacturer Received04/10/2019
Supplement Dates FDA Received04/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-