• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100 A
Device Problem Pressure Problem (3012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/06/2018
Event Type  Injury  
Manufacturer Narrative
Results of investigation: this unit was field serviced by a vyaire medical authorized service technician.The service technician was not able to duplicate the customer's reported issue during testing and evaluation.The ventilator was opened for troubleshooting.Once turned on, the ventilator's oscillator ran for over 45 minutes without shutting down.Adjustments were made to the ddi board, pr3, and pr6.The power supply was checked along with the loop gain pot; all were within specification.The service technician also performed pt calibrations and performance check, all testings were passed per 3100a label.
 
Event Description
It was reported to vyaire medical that the displayed map went to 0.0.There was no patient compromise during this event, the patient was placed on another ventilator with no issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
22745 savi ranch pkwy
yorba linda CA 92887
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
stanley tan
22745 savi ranch pkwy
yorba linda, CA 92887
7149193324
MDR Report Key8201006
MDR Text Key131582168
Report Number2021710-2018-09725
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100 A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/06/2018
Initial Date FDA Received12/27/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-