• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN AP FEM CUTTING BLK SZ 4; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN AP FEM CUTTING BLK SZ 4; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 71434409
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/05/2018
Event Type  malfunction  
Event Description
It was reported that during surgery the ball bearings that hold the valves collett in place came loose.Device broke in patient, all pieces accounted for.No delay or injury reported.No back up required, device use had been completed.
 
Manufacturer Narrative
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.Both ball plungers disassembled from the device and only one was returned with the device.The device was manufactured in 2017 and exhibits signs of moderate wear/usage.A dimensional inspection was performed and an out of tolerance feature was found that could have contributed to the failure.Corrective action is being addressed by the internal quality hold process.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN AP FEM CUTTING BLK SZ 4
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8201038
MDR Text Key131586622
Report Number1020279-2018-02929
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010549174
UDI-Public03596010549174
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71434409
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/05/2018
Initial Date FDA Received12/27/2018
Supplement Dates Manufacturer Received12/05/2018
Supplement Dates FDA Received04/01/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-