• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. ACHIEVE MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. ACHIEVE MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number 2ACH20
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Cardiac Perforation (2513); Pericardial Effusion (3271)
Event Date 11/07/2018
Event Type  Injury  
Manufacturer Narrative
Product event summary: the data files were returned and analyzed.The data files showed at least 7 applications were performed with balloon catheter 2af284 with lot number 85292, without any issue on the date of the event.Clinical issues were encountered during the procedure.There is no indication of a relation of the adverse events to the performance of the physical product.The mapping catheter was not returned for investigation.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryoablation procedure, vital signs were not stable and an epicardial effusion was observed by an echocardiogram.A pericardial puncture was performed to drain the effusion.It was further reported that bleeding could not be stopped, and the patient was transferred to the cardiac department of another facility.The case was aborted and it was noted that the patient was not under general anesthesia.The physician noted that a perforation occurred before the balloon catheter was inserted.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE MAPPING CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key8202008
MDR Text Key131597495
Report Number2182208-2018-02339
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K162892
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/19/2020
Device Model Number2ACH20
Device Catalogue Number2ACH20
Device Lot Number11023564
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/21/2018
Initial Date FDA Received12/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/20/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
2AF284 BALLOON CATHETER, 4FC12 SHEATH
Patient Outcome(s) Life Threatening; Required Intervention;
-
-