Litigation alleges that patient's implant has released metal ions and particles into patient's system, causing severe pain, discomfort, and inflammation in thigh and groin, as well as a popping and snapping sensation in the hip joint when walking or moving to and from a sitting position.Update ad 06 jul 2018 the receipt of ppf and sticker sheets.In addition to what were previously alleged, ppf alleges metal wear and metallosis.
|
Product complaint # (b)(4).Investigation summary = > no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|