• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN LAPRO-CLIP; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN LAPRO-CLIP; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 8886848882
Device Problem Device Slipped (1584)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/30/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during laparoscopic gastrectomy, when the clips were applied to the blood vessel, they came off.A new device was opened with no problems occurred, and the procedure was continued.There was no patient injury.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.He visual inspection noted the cartridge was received applied with the clip body disengaged and the tracks were not inserted properly at the distal end of the clip body.Functionally; the clip body and tracks were reloaded into the cartridge and fired but the clip ejected from the cartridge.The cartridge fingers that hold the clip into place were bent outward due to the pusher opening up the finger after firing.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.A definitive root cause could not be determined with regard to the reported condition because of the condition of the cartridge.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAPRO-CLIP
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key8202699
MDR Text Key131606411
Report Number1219930-2018-06907
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
PMA/PMN Number
K925602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,user faci
Type of Report Initial,Followup
Report Date 03/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2023
Device Model Number8886848882
Device Catalogue Number8886848882
Device Lot NumberN8H0657X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2018
Initial Date Manufacturer Received 12/04/2018
Initial Date FDA Received12/28/2018
Supplement Dates Manufacturer Received02/08/2019
Supplement Dates FDA Received03/04/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-