• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 PINN STRAIGHT CUP IMPACTOR; HIP INSTRUMENTS : IMPACTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 PINN STRAIGHT CUP IMPACTOR; HIP INSTRUMENTS : IMPACTORS Back to Search Results
Catalog Number 221750041
Device Problems Break (1069); Loose or Intermittent Connection (1371); Difficult to Remove (1528); Connection Problem (2900); Device-Device Incompatibility (2919); Material Twisted/Bent (2981); Scratched Material (3020)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/03/2018
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
During a total hip replacement the cup impactor was having difficulty thrading onto the pinnicle cup implant.During the 6.5 screw insertion into the acetabulum the hex screwdriver broke apart inside the patient.Later during an inspection of instrumentation an attune spacer block was found with a chip off one of the bearing surfaces and a rotating platform sz4 articulating surface was found with a crack on the underside.
 
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
No threads are damaged as none of the instrumentation has threading.The chips found from the spacer block was never found, the chip is very small and no damage was noticed during a case with a patient.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: the device associated with this report was not returned.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: the instrument has been received for examination.Examination of the returned instrument confirmed the reported observation.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN STRAIGHT CUP IMPACTOR
Type of Device
HIP INSTRUMENTS : IMPACTORS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8202888
MDR Text Key131620056
Report Number1818910-2018-79378
Device Sequence Number1
Product Code MRA
UDI-Device Identifier10603295098980
UDI-Public10603295098980
Combination Product (y/n)N
PMA/PMN Number
PO70026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 12/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number221750041
Device Lot NumberNT1008
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2019
Initial Date Manufacturer Received 12/03/2018
Initial Date FDA Received12/28/2018
Supplement Dates Manufacturer Received12/30/2018
01/28/2019
02/14/2019
06/05/2019
Supplement Dates FDA Received01/04/2019
02/13/2019
02/15/2019
06/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-