Product event summary: visual inspection of mapping catheter, 990063-020 / 214864046, showed the shaft was kinked and broken at the proximal end attachment with the ecg connector; no ecg signal wires were broken inside the shaft.The product analysis findings do not indicate a manufacturing related defect.In conclusion, the reported broken catheter issue has been confirmed through testing.The mapping catheter failed the returned product inspection due to kinked and broken shaft at ecg connector attachment.If information is provided in the future, a supplemental report will be issued.
|