• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROMODULATION INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROMODULATION INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 3058
Device Problems Human-Device Interface Problem (2949); Insufficient Information (3190)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/19/2018
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a manufacturing representative (rep) regarding a patient implanted with a neurostimulator (ins).The rep reported that they were called in to a case for a lead revision on (b)(6).When the healthcare professional (hcp) opened the pocket, the lead slid out of the ins.The hcp used another ins and connected to the lead already there, impedance tested fine and the hcp closed the pocket.The patient was getting good therapy after.No further complications were reported or are anticipated.
 
Manufacturer Narrative
Analysis of the implantable neurostimulator (ins) model 3058, serial # (b)(4), showed no significant anomalies.It was reported that the set screw was backed out too far.It was noted that the device was subjected to a series of standard tests that included (but was limited to) visual inspection, output, and telemetry.Good, stable out was seen on all electrode pairs.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported later that rep did not know patient's weight and device was returned.
 
Manufacturer Narrative
Analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM II
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
MDR Report Key8203117
MDR Text Key131619131
Report Number3007566237-2018-03727
Device Sequence Number1
Product Code EZW
Combination Product (y/n)N
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 03/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/14/2018
Device Model Number3058
Device Catalogue Number3058
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2019
Initial Date Manufacturer Received 12/21/2018
Initial Date FDA Received12/28/2018
Supplement Dates Manufacturer Received01/02/2019
03/26/2019
Supplement Dates FDA Received01/10/2019
03/27/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age36 YR
-
-