• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE; UTETERAL STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE; UTETERAL STENT Back to Search Results
Model Number 777626
Device Problems Device Damaged Prior to Use (2284); Material Split, Cut or Torn (4008)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 12/04/2018
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the ureteral stent had a split in the tip, this was found when the package was opened.
 
Event Description
It was reported that the ureteral stent had a split in the tip, this was found when the package was opened.
 
Manufacturer Narrative
The reported event was unconfirmed.A ureteral stent was returned.There were no visual defects and neither ends of the stent were split.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿precautions: for single use only.Do not resterilize.Do not use if the package or product is damaged.Improper handling technique can seriously weaken the stent.Acute bending or overstressing during placement could result in subsequent separation of the stent at the point of stress after a prolonged indwelling period.Suture may be cut off prior to stent placement.Remove suture prior to placement for pediatric patients.Exercise care.Tearing of the stent can be caused by sharp instruments.Ureteral stents should be checked periodically for signs of encrustation and proper function.Periodic checks of the stent by cystoscopic and/or radiographic procedures are recommended at intervals deemed to be appropriate by the physician in consideration of the individual patient¿s condition and other patient specific factors.When long-term use is indicated, it is recommended that indwelling time not exceed 365 days.The stent is not intended as a permanent indwelling device.Care should be exercised when removing the stent from inner polybag so as not to cause tearing or fragmentation.With any ureteral stent, migration is a possible complication which could require medical intervention for removal.Selection of too short a stent may result in migration.The insertion of a ureteral stent should only be done by those individuals who have comprehensive training in the techniques and risks of the procedure.Multi-length ureteral stents: formation of knots in multi-length ureteral stents may occur.This may result in injury to the ureter during removal and/or the need for additional surgical intervention.The presence of a knot should be considered if significant resistance is encountered during attempts at removal.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE
Type of Device
UTETERAL STENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8204035
MDR Text Key131708237
Report Number1018233-2018-06250
Device Sequence Number1
Product Code FAD
UDI-Device Identifier10801741014557
UDI-Public(01)10801741014557
Combination Product (y/n)N
PMA/PMN Number
K983498
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup
Report Date 06/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/28/2022
Device Model Number777626
Device Catalogue Number777626
Device Lot NumberNGCP1885
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2019
Initial Date Manufacturer Received 12/04/2018
Initial Date FDA Received12/28/2018
Supplement Dates Manufacturer Received05/15/2019
Supplement Dates FDA Received06/04/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-