ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME TEST STRIPS Back to Search Results
Catalog Number 04625315160
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/22/2018
Event Type  malfunction  
Manufacturer Narrative
[mw5081790].
 
Event Description
This event was reported by the customer to fda as mw5081790.The customer complained of discrepant inr results for 1 patient tested on a coaguchek xs meter with an unspecified serial number compared to an unspecified laboratory method.The date of event is an approximation as the date of event was not provided.The result from the meter was > 6 inr.The actual result was not provided.The result from the laboratory was "in the 3 range" inr.The actual result was not provided.There was no allegation that an adverse event occurred.The complained test strips have been calibrated against the who standard and are in scope of the roche initiated recall.For these test strips there is a potential for a product problem when the inr is > 4.5.Values > 4.5 inr showed an increasing positive bias.The information in the case is consistent with the details of the recall and the issue has been fully investigated.
 
Manufacturer Narrative
Coaguchek uses human recombinant thromboplastin.Therefore, comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key8204054
MDR Text Key131772357
Report Number1823260-2018-05171
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 04/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04625315160
Device Lot NumberASKU
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/17/2018
Initial Date FDA Received12/28/2018
Supplement Dates Manufacturer Received12/17/2018
Supplement Dates FDA Received04/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberRES Z-0360-2019
Patient Sequence Number1
Treatment
WARFARIN SODIUM
-
-