• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ALLURE RF; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ALLURE RF; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3222
Device Problem Pacemaker Found in Back-Up Mode (1440)
Patient Problem No Patient Involvement (2645)
Event Date 12/19/2018
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that prior to the procedure, the in-box pulse generator exhibited backup operation.The device was not used.
 
Manufacturer Narrative
The reported event of backup operation was confirmed.Analysis of the device image determined backup occurred prior to distribution due to telemetry commanded microprocessor reset, consistent with unintended programmer interaction.Device firmware was successfully reloaded, and further environmental testing could not reproduce the reset.Device battery was found to be above elective replacement level.Longevity assessment was performed, and device was in the normal range of operation with appropriate remaining longevity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURE RF
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8204197
MDR Text Key131701537
Report Number2017865-2018-19743
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734507905
UDI-Public05414734507905
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2018
Device Model NumberPM3222
Device Lot NumberA000041175
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/19/2018
Initial Date FDA Received12/28/2018
Supplement Dates Manufacturer Received05/13/2019
Supplement Dates FDA Received05/14/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/07/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-