• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VICMO12.1
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Code Available (3191)
Event Date 10/25/2018
Event Type  Injury  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.(b)(4).
 
Event Description
The reporter indicated the surgeon implanted a 12.1mm vicmo12.1 implantable collamer lens, -06.50 diopter, in the patient's right eye (od) on (b)(6) 2018.The lens was explanted on (b)(6) 2018 due to low vaulting.The lens was exchanged for a longer lens and the problem was resolved.The reporter indicated the cause of the event was unknown.
 
Manufacturer Narrative
Device evaluation: lens was returned dry in a micro-centrifuge vial.There was clear surgical residue on product.Visual inspection found the haptic bent and presence of residue on lens surface.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key8204853
MDR Text Key131696547
Report Number2023826-2018-01933
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 12/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/30/2020
Device Model NumberVICMO12.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2019
Initial Date Manufacturer Received 12/03/2018
Initial Date FDA Received12/28/2018
Supplement Dates Manufacturer Received02/08/2019
Supplement Dates FDA Received03/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age20 YR
-
-