• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD AIRVO 2 HUMIDIFIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD AIRVO 2 HUMIDIFIER Back to Search Results
Model Number PT101
Device Problem No Audible Alarm (1019)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint airvo humidifier was received at our fisher & paykel healthcare (f&p) regional office in (b)(6) and was inspected by a trained f&p technician.The device was performance tested and the audible alarm function was checked.The device was then disposed of.Results: during testing it was found that the visual alerts on the complaint airvo were operating, however the audible alarm did not function.Previous investigations into this type of failure have identified that the problem is caused by a faulty speaker and electrical resistance testing has shown the speaker's resistance to be open circuit.Conclusion: as part of our ongoing product improvement initiatives, we have implemented a soak test for 100% testing of the speaker harness on the airvo production line, which identifies and discards any faulty speakers prior to assembly into the airvo.Additionally, a new speaker unit has more recently been sourced from a different supplier.The subject airvo was manufactured prior to implementation of these measures.The airvo user manual states that the "airvo is for the treatment of spontaneously breathing patients who would benefit from receiving high-flow, warmed and humidified respiratory gases" and that "the unit is not intended for life support." the user manual warns the user: prior to each patient use, ensure that the auditory alarm signal is audible by conducting the alarm system functionality check described in the alarms section.The alarm system functionality check instructs the user on how to test the alarm and states that "if either alarm signal is absent, do not use the unit.Contact your fisher & paykel healthcare representative.".
 
Event Description
A hospital in (b)(6) reported a fault with an pt101 airvo 2 humidifier.Upon assessment of the device at the fisher & paykel healthcare (f&p) regional office on (b)(6) 2018, it was found that the audible alarm was not working properly.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRVO 2 HUMIDIFIER
Type of Device
AIRVO 2 HUMIDIFIER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, CA 92618
9494534000
MDR Report Key8205407
MDR Text Key131944564
Report Number9611451-2018-01138
Device Sequence Number1
Product Code BTT
UDI-Device Identifier09420012422354
UDI-Public(01)09420012422354
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K131895
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPT101
Device Catalogue NumberPT101
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/01/2018
Initial Date FDA Received12/30/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/17/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-