• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VTICMO13.2
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Code Available (3191)
Event Date 08/10/2018
Event Type  Injury  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.(b)(4).
 
Event Description
The reporter indicated the surgeon implanted a 13.2mm vticmo13.2 implantable collamer lens, -15.0/+1.5/083 (sphere/cylinder/axis), in the patients left eye (os), on (b)(6) 2018.The lens was explanted on (b)(6) 2018 due to excessive vaulting.The lens was exchanged for a shorter lens and the problem was resolved.
 
Manufacturer Narrative
Lens was returned in a micro-centrifuge vial with moisture on lens surface.Visual inspection found no visible damage to lens.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key8207006
MDR Text Key131741688
Report Number2023826-2018-01942
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2019
Device Model NumberVTICMO13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/03/2018
Initial Date FDA Received12/31/2018
Supplement Dates Manufacturer Received02/08/2019
Supplement Dates FDA Received03/05/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/17/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age40 YR
-
-