• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CASHEL 466P306X; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CASHEL 466P306X; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number 466P306X
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem Perforation of Vessels (2135)
Event Date 12/06/2018
Event Type  Injury  
Manufacturer Narrative
As reported by the legal brief, the patient underwent placement of an trapease vena cava filter.The filter subsequently malfunctioned and caused injury and damages to the patient including, but not limited to, filter perforation - mesenteric.
 
Event Description
As reported by the legal brief, the patient underwent placement of an trapease vena cava filter.The filter subsequently malfunctioned and caused injury and damages to the patient including, but not limited to, filter perforation - mesenteric.
 
Manufacturer Narrative
It was reported that a patient underwent placement of a trapease vena cava filter.The information provided indicated that the filter subsequently malfunctioned and caused injury and damages to the patient including, but not limited to, filter perforation - mesenteric.There is currently no additional information available for review.The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.The trapease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Vessel perforation is a known adverse event associated with implanting vena cava filters and is listed as such in the instructions for use (ifu).The ifu also notes vessel damage such as intimal tears and perforation as procedural and long-term complications related to ivc filters.With the limited information available and without the procedural films or post implant images to review the reported device perforation could not be confirmed or further clarified.Given the limited information currently available for review, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
466P306X
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
CORDIS CASHEL
cahir road
cashel, co. tipperary
EI 
MDR Report Key8207823
MDR Text Key131768935
Report Number1016427-2019-02373
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
PMA/PMN Number
K020316
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup
Report Date 01/31/2019
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number466P306X
Device Catalogue Number466P306X
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/06/2018
Initial Date FDA Received01/02/2019
Supplement Dates Manufacturer Received01/05/2019
Supplement Dates FDA Received01/31/2019
Patient Sequence Number1
-
-