• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 OFFSET IMPACTOR; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 OFFSET IMPACTOR; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Catalog Number 71368569
Device Problems Degraded (1153); Unintended Movement (3026); Appropriate Term/Code Not Available (3191)
Patient Problems Injury (2348); No Information (3190)
Event Date 12/05/2018
Event Type  malfunction  
Event Description
It was reported that, during surgery, device was not working properly.No back-up was available.
 
Manufacturer Narrative
The associated complaint device was returned and evaluated.A visual inspection of the returned r3 offset impactor indicated that the thumb release no longer holds.A functional evaluation indicated that the linkage does not function smoothly.The plastic impactor tip is missing.Failure to use this tip may result in the distal linkage section to become damaged and the overall device not functioning as intended.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history on the listed part revealed no prior complaints for the listed batch.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 OFFSET IMPACTOR
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8210199
MDR Text Key131851381
Report Number1020279-2019-00016
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010600240
UDI-Public03596010600240
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71368569
Device Lot Number18DM06021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/31/2018
Initial Date Manufacturer Received 12/05/2018
Initial Date FDA Received01/02/2019
Supplement Dates Manufacturer Received12/05/2018
Supplement Dates FDA Received01/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-